{{DISPLAYTITLE:Calibrate your the scanner|noerror}}
<translate>
<!--T:4-->
In this page , it will be explained how to calibrate the scanner.
== Recalibrate Head== <!--T:5-->
#Put the calibration kit inside the scanner in the correct position
#Select the calibration type from the drop-down menu
#In the calibration dialog, select from the drop -down menu the exact calibration kit name
#Click '''Start'''
<!--T:11-->
If the calibration kit name is not available in the drop -down menu, the '''Load New Master XmlXML''' button can be used to add a new calibration kit from the appropriate Xml XML description file.
<!--T:12-->
This tool will start automatically after a '''Head Recalibration''' or can be started manually with the {{Inline button|axisCalibAction.png}} button.
<!--T:31-->
The software will prompt the user to insert the reference rim on the model holder
<!--T:32-->
'''NOTE THAT THE REFERENCE RIM IS A PRECISION OBJECT! OCCLUSAL AND WHITENING SPRAYS, AND CLEANING SOLVENTS, DAMAGE ITS SURFACE!'''
'''DO NOT SPRAY ANY SUBSTANCE ON THE RIM AS TO DO SO IT WILL NO LONGER BE USABLE FOR SCANNING NOR FOR CALIBRATING!'''
<!--T:33-->
[[File:service-rim-message.png]]
<!--T:19-->
<!--T:24-->
*'Axes Recalibration Successfull' : the calibration has been completed,
*'Axes Recalibration Successfull, Dynamic Articulation Calibration Failed' : the calibration has been completed, however , the dynamic articulation positioning will not work. This message shows when the dynamic articulation module is active on the license, but the scanner has been calibrated with a normal stone model.*'Axes Recalibration Failed' : calibration has to be repeated
==Color Calibration== <!--T:25-->
<!--T:26-->
This last calibration, that can be opened by clicking on the icon {{Inline button|colorCalibAction.png}}, allows to calibrate calibrating the acquired colors of the scanner to have realistic color results.